Revision 1.3 P/N 119-042 9 January 2008
Lake Shore Cryotronics, Inc.
575 McCorkle Blvd.
Westerville, Ohio 43082-8888 USA
E-mail Addresses:
Visit Our Website At:
www.lakeshore.com
Fax: (614) 891-1392
Telephone: (614) 891-2243
Methods and apparatus disclosed and described herein have been developed solely on company funds of Lake Shore Cryotronics, Inc. No government
or other contractual support or relationship whatsoever has existed which in any way affects or mitigates proprietary rights of Lake Shore Cryotronics,
Inc. in these developments. Methods and apparatus disclosed herein may be subject to U.S. Patents existing or applied for. Lake Shore Cryotronics,
Inc. reserves the right to add, improve, modify, or withdraw functions, design modifications, or products at any time without notice. Lake Shore shall
not be liable for errors contained herein or for incidental or consequential damages in connection with furnishing, performance, or use of this material.
User’s Manual
Model 642
Electromagnet
Power Su
pp
l
y
Przeglądanie stron 0
1 2 3 4 5 6 ... 121 122

Podsumowanie treści

Strona 1 - Model 642

Revision 1.3 P/N 119-042 9 January 2008 Lake Shore Cryotronics, Inc. 575 McCorkle Blvd. Westerville, Ohio

Strona 2

Lake Shore Model 642 Electromagnet Power Supply User’s Manual x Table of Contents LIST OF TABLES Table No. Title Page 3-1 Rear Panel Connecto

Strona 3

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 7-8 Service Firmware Replacement (Continued) 4. Disconnect the ribbon cable to the

Strona 4

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Service 7-9 DIGITAL BOARD Figure 7-5. Board Locations (top view)

Strona 5 - TABLE OF CONTENTS

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 7-10 Service VR1R74C67CR18L1AUXILIARYC131K3CR10R73K2VR3K5C132J10J9MAGNETF9C130R71V

Strona 6

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Service 7-11 7.10 CONNECTOR AND CABLE DEFINITIONS All non-power electrical connectio

Strona 7

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 7-12 Service 7.10.3 Auxiliary Connector The Auxiliary Connector provides connection

Strona 8

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Service 7-13 7.10.4 Power Supply Connector The Power Supply connector provides the

Strona 9 - LIST OF ILLUSTRATIONS

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 7-14 Service 7.10.6 Serial Interface Cable Wiring The following are suggested cabl

Strona 10 - LIST OF TABLES

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Service 7-15 1212413 Pin Symbol Description 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 1

Strona 11 - INTRODUCTION

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 7-16 Service 7.11 CALIBRATION Lake Shore maintains a fully automated calibration fi

Strona 12 - 1-2 Introduction

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Service 7-17 7.11.2 Calibration Equipment 1. Calibration Shunt Resistor – The outpu

Strona 13 - 1.2 SPECIFICATIONS

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Introduction 1-1 CHAPTER 1 INTRODUCTION 1.0 GENERAL This chapter provides an intro

Strona 14 - Specifications (Continued)

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 7-18 Service 7.13.3.3 Calibrate Output Voltage Reading Zero This step assumes that

Strona 15 - Introduction 1-5

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Service 7-19 Calibrate Output Current Gain (Continued) 12. Measure the actual volta

Strona 16 - 1-6 Introduction

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 7-20 Service 7.13.3.6 Calibrate Current Reading Gain 1. Send “CALG 0, 1”. To set

Strona 17 - 1.3 SAFETY SUMMARY

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Service 7-21 7.11.3.8 Calibrate External Current Programming Voltage Reading Gain T

Strona 18 - Safety Summary (Continued)

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 7-22 Service CALG? Gain Calibration Constant Query Input: CALG? <type>[te

Strona 19 - 1.4 SAFETY SYMBOLS

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Glossary of Terminology A-1 APPENDIX A GLOSSARY OF TERMINOLOGY accuracy. The degre

Strona 20 - 1-10 Introduction

Lake Shore Model 642 Electromagnet Power Supply User’s Manual A-2 Glossary of Terminology boiling point. The temperature at which a substance in

Strona 21 - INSTALLATION, AND OPERATION

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Glossary of Terminology A-3 four-lead. measurement technique where one pair of excit

Strona 22

Lake Shore Model 642 Electromagnet Power Supply User’s Manual A-4 Glossary of Terminology Kelvin Scale. The Kelvin Thermodynamic Temperature Scal

Strona 23 - Magnet System Design 2-3

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Glossary of Terminology A-5 permeability. Material parameter which is the ratio of t

Strona 24

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 1-2 Introduction Output Architecture (Continued) The Model 642 output architecture

Strona 25

Lake Shore Model 642 Electromagnet Power Supply User’s Manual A-6 Glossary of Terminology Seebeck effect. The development of a voltage due to dif

Strona 26 - 2-6 Magnet System Design

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Units for Magnetic Properties B-1 APPENDIX B UNITS FOR MAGNETIC PROPERTIES Table B

Strona 27 - Magnet System Design 2-7

Lake Shore Model 642 Electromagnet Power Supply User’s Manual B-2 Units for Magnetic Properties Table B-2. Recommended SI Values for Physical C

Strona 28 - 2-8 Magnet System Design

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Introduction 1-3 Description (Continued) Interfaces The Model 642 includes IEEE-488

Strona 29 - INSTALLATION

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 1-4 Introduction Specifications (Continued) Output programming Internal current set

Strona 30 - 3.2 REAR PANEL DEFINITION

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Introduction 1-5 Specifications (Continued) Interface IEEE-488.2 interface Features

Strona 31

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 1-6 Introduction Specifications (Continued) General Line power Power 5500 VA max Vo

Strona 32 - VOLTAGE INDICATOR KEY

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Introduction 1-7 Accessories included MAN-642 Model 642 user manual 6031 Two front

Strona 33

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 1-8 Introduction Safety Summary (Continued) Do Not Substitute Parts Or Modify Instr

Strona 34 - 3.3.6 Wiring Cover

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Introduction 1-9 Safety Summary (Continued) 1.4 SAFETY SYMBOLS NumberSymbol12345678

Strona 35 - TO REMOTE ALARM CIRCUIT

Lake Shore Model 642 Electromagnet Power Supply User’s Manual ii LIMITED WARRANTY STATEMENT WARRANTY PERIOD: ONE (1) YEAR 1. Lake Shore warran

Strona 36 - VALVE AND SWITCH

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 1-10 Introduction This Page Intentionally Left Blank

Strona 37 - 230 mm (9 in) LONG

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Magnet System Design 2-1 CHAPTER 2 MAGNET SYSTEM DESIGN, INSTALLATION, AND OPERATIO

Strona 38 - LUG COVER

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 2-2 Magnet System Design 2.2 MAGNET CONSTRUCTION The magnet consists of two water-c

Strona 39

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Magnet System Design 2-3 POWER LEADPOWER LEADCROSS CONNECTION WIRINGPOWER LEADPOWER

Strona 40 - 3.13 RACK MOUNTING

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 2-4 Magnet System Design 2.3.3 Temperature Switches and Flow Switches As discussed

Strona 41 - OPERATION

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Magnet System Design 2-5 2.3.6 Final Check-Out When all of the connections have been

Strona 42 - +70.0000 A +35.000 V

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 2-6 Magnet System Design Figure 2-5. Typical Curves of Field vs. Current for Va

Strona 43 - 4.4 KEYPAD DEFINITION

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Magnet System Design 2-7 Figure 2-5. Typical Curves of Field vs. Current for Vario

Strona 44 - 4.6 SETTING OUTPUT CURRENT

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 2-8 Magnet System Design 2.4.4 Avoiding Cooling Water Condensation If the temperat

Strona 45 - 4.8 RAMP SEGMENTS

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Installation 3-1 CHAPTER 3 INSTALLATION 3.0 GENERAL This chapter provides general

Strona 46 - 4.10 ZERO OUTPUT

Lake Shore Model 642 Electromagnet Power Supply User’s Manual iii

Strona 47 - 4.11 MAXIMUM SETTING LIMITS

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 3-2 Installation 3.2 REAR PANEL DEFINITION This paragraph defines the rear panel of

Strona 48 - 4.14 ERROR STATUS DISPLAY

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Installation 3-3 12312876101591413151154 Figure 3-1. Model 642 Rear Panel (shown wi

Strona 49 - 4.16 LOCKING THE KEYPAD

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 3-4 Installation Line Voltage Selection (Continued) Changing line voltage is accomp

Strona 50 - 4.17 COMPUTER INTERFACE

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Installation 3-5 3.3.3 Start-up Fuses The start-up transformer and associated circui

Strona 51

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 3-6 Installation POWER CABLESTRIP 13mm (0.5 in)WIRE FERRULE (OPTIONAL)POWER TERMINA

Strona 52

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Installation 3-7 3.3.7 Mains Wiring No cord or disconnecting plug is provided with t

Strona 53 - COMPUTER INTERFACE OPERATION

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 3-8 Installation 3.6 POWER SUPPLY CONNECTOR The Power Supply connector provides te

Strona 54 - 5.1.2 Remote/Local Operation

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Installation 3-9 SOLENOID VALVEWITH MOUNTING BRACKETADJUSTABLE HOSE CLAMPREINFORCED

Strona 55

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 3-10 Installation BELLEVILLE WASHERBOLTOUTPUT LUGMAGNET CABLEPLAIN WASHERNUT Figure

Strona 56 - 5.1.4.1 Overview

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Installation 3-11 3.9 ANALOG INPUT/OUTPUT CONNECTIONS The Analog I/O connector prov

Strona 57

Lake Shore Model 642 Electromagnet Power Supply User’s Manual iv This Page Intentionally Left Blank

Strona 58

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 3-12 Installation 3.12 DETACHABLE HANDLES The Model 642 is supplied with 4 detachab

Strona 59 - Register Method Example

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Operation 4-1 CHAPTER 4 OPERATION 4.0 GENERAL This chapter provides operating inst

Strona 60 - *ESE, *ESE?

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 4-2 Operation 4.2 DISPLAY DEFINITION The Model 642 has an 8 line by 40 character va

Strona 61

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Operation 4-3 4.4 KEYPAD DEFINITION The Model 642 has 22 keys separated into 3 group

Strona 62 - ERSTE, ERSTE?

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 4-4 Operation 4.4.1 General Keypad Operation The Model 642 uses three basic keypad

Strona 63

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Operation 4-5 Setting Output Current (Continued) To change the output current settin

Strona 64 - *SRE, *SRE?

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 4-6 Operation Ramp Segments (Continued) To configure the ramp segments press Ramp S

Strona 65

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Operation 4-7 4.11 MAXIMUM SETTING LIMITS The Model 642 offers a maximum setting lim

Strona 66

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 4-8 Operation 4.12 MAGNET WATER The Model 642 provides power to control an external

Strona 67

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Operation 4-9 4.15 EXTERNAL CURRENT PROGRAMMING The output current of the Model 642

Strona 68

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Table of Contents v TABLE OF CONTENTS Chapter/Section Title Page 1 INTRODUCTION ..

Strona 69

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 4-10 Operation Locking the Keypad (Continued) To lock or unlock the instrument keyp

Strona 70

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Operation 4-11 4.17.2 Changing IEEE-488 Interface Parameters Two interface paramete

Strona 71 - IBCONF.EXE.eps

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 4-12 Operation Table 4-3. Default Parameter Values Output Settings Output Current

Strona 72 - 5.1.6 Troubleshooting

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-1 CHAPTER 5 COMPUTER INTERFACE OPERATION 5.0 GENERA

Strona 73 - 5.2.2 Physical Connection

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-2 Computer Interface Operation 5.1.1 Changing IEEE-488 Interface Parameters Two

Strona 74

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-3 5.1.3.1 Bus Control Commands A Universal Command a

Strona 75 - 5.2.6 Message Flow Control

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-4 Computer Interface Operation Message Strings (Continued) A query string is issu

Strona 76

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-5 Service RequestNameBitNotRQS = Service RequestMSS =

Strona 77

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-6 Computer Interface Operation NameNameNameUsedNotUsedNot76543210BitUsed UsedERS

Strona 78

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-7 5.1.4.1.6 Reading Registers Any register in the sta

Strona 79 - 5.2.8 Troubleshooting

Lake Shore Model 642 Electromagnet Power Supply User’s Manual vi Table of Contents TABLE OF CONTENTS (Continued) Chapter/Section Title Page 4 O

Strona 80 - 5.3 COMMAND SUMMARY

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-8 Computer Interface Operation 5.1.4.2 Status Register Sets As shown in Figure 5

Strona 81

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-9 5.1.4.2.2 Operation Event Register Set The Operati

Strona 82

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-10 Computer Interface Operation 5.1.4.3 Error Status Register Sets As shown in F

Strona 83

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-11 5.1.4.3.2 Operational Error Status Register Set T

Strona 84

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-12 Computer Interface Operation 5.1.4.4 Status Byte and Service Request (SRQ) As

Strona 85

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-13 5.1.4.4.2 Service Request Enable Register The Ser

Strona 86

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-14 Computer Interface Operation 5.1.4.4.6 Using Operation Complete (*OPC) and Op

Strona 87

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-15 Figure 5-7. GPIB0 Setting Configuration Figur

Strona 88

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-16 Computer Interface Operation 5.1.5.2 Visual Basic IEEE-488 Interface Program S

Strona 89

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-17 Table 5-4. IEEE-488 Interface Program Control Prop

Strona 90

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Table of Contents vii TABLE OF CONTENTS (Continued) Chapter/Section Title Page 5.1

Strona 91 - OPTIONS AND ACCESSORIES

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-18 Computer Interface Operation Table 5-5. Visual Basic IEEE-488 Interface Progr

Strona 92 - 6-2 Options and Accessories

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-19 Primary GPIB Address ... Primary GPIB Addres

Strona 93 - SERVICE

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-20 Computer Interface Operation 5.1.5.3 Program Operation Once the running, try t

Strona 94 - 7.3 LINE VOLTAGE SELECTION

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-21 5.2 SERIAL INTERFACE OVERVIEW The serial interfac

Strona 95

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-22 Computer Interface Operation 5.2.3 Hardware Support The Model 642 interface ha

Strona 96 - Figure 7-3. Circuit Breaker

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-23 Message Strings (Continued) A query string is issu

Strona 97 - 7.6 ERROR MESSAGES

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-24 Computer Interface Operation 5.2.7.1 Visual Basic Serial Interface Program Set

Strona 98 - 7.7 ELECTROSTATIC DISCHARGE

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-25 Table 5-7 Serial Interface Program Control Proper

Strona 99

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-26 Computer Interface Operation Table 5-8. Visual Basic Serial Interface Program

Strona 100 - 7-8 Service

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-27 5.2.7.2 Program Operation Once the example program

Strona 101 - DIGITAL BOARD

Lake Shore Model 642 Electromagnet Power Supply User’s Manual viii Table of Contents TABLE OF CONTENTS (Continued) Chapter/Section Title Page 7

Strona 102 - MOUNTING SCREWS

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-28 Computer Interface Operation 5.3 COMMAND SUMMARY This paragraph provides a lis

Strona 103

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-29 Table 5-9. Command Summary Command Function Page C

Strona 104 - Pin Name

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-30 Computer Interface Operation *ESE? Standard Event Status Enable Register Que

Strona 105

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-31 *SRE Service Request Enable Register Command Inp

Strona 106 - 7-14 Service

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-32 Computer Interface Operation DFLT Factory Defaults Command Input: DFLT 99[ter

Strona 107 - Pin Symbol Description

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-33 ERSTR? Error Status Register Query Input: ERSTR?

Strona 108 - 7.11.1 Calibration Interface

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-34 Computer Interface Operation LIMIT Limit Output Settings Command Input: LIM

Strona 109

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-35 OPST? Operational Status Query Input: OPST? [ter

Strona 110 - 7-18 Service

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-36 Computer Interface Operation RDGV? Output Voltage Reading Query Input: RDGV?

Strona 111 - Service 7-19

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Computer Interface Operation 5-37 STOP Stop Output Current Ramp Command Input: STO

Strona 112 - 7-20 Service

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Table of Contents ix LIST OF ILLUSTRATIONS Figure No. Title Page 1-1 Model 642 Fr

Strona 113 - Service 7-21

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 5-38 Computer Interface Operation This Page Intentiona

Strona 114 - 7-22 Service

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Options and Accessories 6-1 CHAPTER 6 OPTIONS AND ACCESSORIES 6.0 GENERAL This cha

Strona 115 - GLOSSARY OF TERMINOLOGY

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 6-2 Options and Accessories This Page Intentionally Le

Strona 116

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Service 7-1 CHAPTER 7 SERVICE 7.0 GENERAL This chapter provides basic service info

Strona 117 - – 1855) a German

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 7-2 Service 7.3 LINE VOLTAGE SELECTION The Model 642 may be configured for four ba

Strona 118

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Service 7-3 TERMINAL SCREWWIRE FERRULEVOLTAGE SELECTOR CARDVOLTAGE INDICATOR KEY Fi

Strona 119 - Factor Prefix Symbol

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 7-4 Service CURRENT SETTING DIALACCESS COVER Figure 7-3. Circuit Breaker 7. If i

Strona 120

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Service 7-5 Figure 7-4. Fuse Holder Detail 7.6 ERROR MESSAGES The following messa

Strona 121 - APPENDIX B

Lake Shore Model 642 Electromagnet Power Supply User’s Manual 7-6 Service Table 7-3. Operational Errors Remote Enable Fault Detected The Remote E

Strona 122

Lake Shore Model 642 Electromagnet Power Supply User’s Manual Service 7-7 7.7.2 Handling Electrostatic Discharge Sensitive Components Observe all